GENERAL DESCRIPTION BIOTEMP Temperature Probes Series is recommended for its compatibility with various kinds of hospital equipment.BIOGENESIS guarantees the interchange and stability, with a temperature margin that is less than 0,1ºC.
WARNING Sensors are susceptible to error when exposed to high radio frequency intensity.The electric bistouries is a source of high radio frequency.
DESINFECTION
The most used is ethylene oxide. Oxygen water at 3vol , isoprophanol at 70%, ethanol at 70%, CIDEX, solution 1:10 of sodium hypochlorite can be used.
WARRANTY
All BIOGENESIS
probes
have a 1 year of production warranty.
The warranty does not cover bad treatments or misuse. The useful life of our sensor depends of its use and treatment.
GENERAL
SPECS
-Temperature Range: 0ºC a 60ºC.
-Accuracy: +
0.1C from 32°C to 42°C + 0.2°C in the rest of the scale.
-Interchangeable: 0.1°C from
0°C to 60°C.
CLEANING
INSTRUCTIONS
Clean the Sensors with an H2O solution and a neutral detergent Do not submerge.
Precautions
Do not use in autoclave (sterilizer).
The sensor can be scratched when pulled, misuse can cause the sensor to
break.
Avoid contact with alcohol, detergent or another alkaline
substance.
If you use with ether, acetone or ester the sensor will lose its
flexibility. Please store sensors in a 25ºC environment. The temperature should not exceed 50ºC.
The Temperature sensor must be used by qualified personnel. |
||
|
|
||
|
|
||
|
|
||